B. CP11410 AUTOREAD MUELLER-HINTON BROTH W/TES W/LYSED HORSE BLOOD
CP112-10
626905
Manufacturer
Oxoid Limited
Wade Road
Basingstoke, Hampshire
RG24 8PW
UNITED KINGDOM
C. CP11410 AUTOREAD MUELLER-HINTON BROTH W/TES W/LYSED HORSE BLOOD
T3462
613260
Manufacturer
Oxoid Limited
Wade Road
Basingstoke, Hampshire
RG24 8PW
UNITED KINGDOM
A. CP11410 AUTOREAD MUELLER-HINTON BROTH W/TES W/LYSED HORSE BLOOD
CP11410
Introduction
Affected Products
A. CP11410 AUTOREAD MUELLER-HINTON BROTH W/TES W/LYSED HORSE BLOOD
B. CP11410 AUTOREAD MUELLER-HINTON BROTH W/TES W/LYSED HORSE BLOOD
C. CP11410 AUTOREAD MUELLER-HINTON BROTH W/TES W/LYSED HORSE BLOOD
Reason
The use of the listed lots for these sensititre broth could produce out of specification high quality control (QC) results (false resistance) for the antibiotic tigecycline with QC organism Enterococcus faecalis. The risk associated with false resistance to the antibiotic would be potential delay in reporting; therefore the risk associated with this incident is considered low.