POWER PROCESSOR SAMPLE PROCESSING SYSTEM (2015-10-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
90HP1301002POWER PROCESSOR SAMPLE PROCESSING SYSTEM
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
92821
UNITED STATES
A11036

Introduction

Affected Products

POWER PROCESSOR SAMPLE PROCESSING SYSTEM (2015-10-05)

Reason

Beckman Coulter has identified that the centrifuge module on the power processor sample processing system exceeds the published acoustic noise specification of 74 decibels. Beckman coulter implemented a new centrifuge motor on the power processor which included an additional cooling fan mounted directly on the motor shaft that spins when the centrifuge motor runs during centrifugation. As a result, centrifuges with the new motor have two cooling fans that are positioned very close to each other and spin in opposite directions causing air turbulence and acoustic noise.