Health Products

MONACO (2015-10-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Version 5.10.00 and higher
Manufacturer
Elekta Business Area Software Systems Impac Medical Systems, Inc.
13723 RIVERPORT DRIVE
MONACOMONACO

Introduction

Affected Products

MONACO

Reason

When the 100% isodose value is edited on the isodose control, the dose normalization is automatically updated. If the user does not notice this update and then rescales to the relative isoline and does not read the dose normalization value displayed on the rescale bar, the rescale could be other than intended. When the user optimizes, then changes the bolus assigment of the beams, and then selects "calculate" (not "optimize" again), the bolus assignment to beams can get scrambled.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime