HEATER-COOLER UNIT 30 SYSTEM - MAIN UNIT (2015-10-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.HEATER-COOLER UNIT 30 SYSTEM - MAIN UNIT
Manufacturer
MAQUET CARDIOPULMONARY GMBH
KEHLER STR. 31, RASTATT, GERMANY, 76437
RASTATT
76437
GERMANY
00705501

Introduction

Affected Products

HEATER-COOLER UNIT 30 SYSTEM - MAIN UNIT

Reason

MAQUET Cardiopulmonary has received complaints associated with a failure of the power supply board of the HCU 30. An internal investigation showed that a drop of more than 10% voltage can cause an overheating and subsequent failing of the NTC (negative temperature coefficient) R56 thermistor of the power supply board. This failure typically occurs when the compressor is switched on, i.e. before patient connection. A defective thermistor results in an inoperable compressor or in an inoperable HCU 30. There are no complaints reported associated with a patient injury or death.