Health Products

GETINGE WASHER/DISINFECTOR (2015-10-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Getinge Disinfection AB
LJUNGADALSGATAN 11
VAXJO
35115
SWEDEN
GETINGE WASHER/DISINFECTOR46-2

Introduction

Affected Products

GETINGE WASHER/DISINFECTOR

Reason

Getinge received two complaints that sediment residuals were found in the manifold of Getinge 46 Series Washer/Disinfectors with the drying option and PACS 300 controls. Investigation revealed that there was insufficient water flow through the manifold, which allowed air pockets to occur.