XPER FLEX CARDIO PHYSIOMONITORING SYSTEM-MAIN UNIT (2015-10-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Invivo, a Div. of Philips Medical Systems
12151 Research Parkway
Orlando
32826
Florida
UNITED STATES
XPER FLEX CARDIO PHYSIOMONITORING SYSTEM-MAIN UNITFC2010

Introduction

Affected Products

XPER FLEX CARDIO PHYSIOMONITORING SYSTEM-MAIN UNIT

Reason

Philips has become aware that the SpO2 (peripheral capillary oxygen saturation) and non-invasive blood pressure (NIBP) waveform and numeric values displayed on the device may freeze, resulting in the display of measurements that are not current. Although SpO2 and NIBP monitoring and alarms are no longer functional, all other monitoring parameters are functional and visible. If this problem were to occur, the device can be reset by power cycling and the reset will allow the customer to continue monitoring. However, power cycling the device does not prevent the issue from recurring.