MR SURGICAL SUITE II TABLE (2015-10-29)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
416369MR3TMR SURGICAL SUITE II TABLE
Manufacturer
GE Medical Systems, LLC
3000 North Grandview Blvd.
Waukesha
53188
Wisconsin
UNITED STATES
M0001SS

Introduction

Affected Products

MR SURGICAL SUITE II TABLE

Reason

Potential safety issue with the table transfer release of the MR SURGICAL SUITE II MR System. The issue may cause an unexpected function loss of the patient transfer mechanism that allows the patient to be moved between the MR and surgical tables. This could delay care and handling of a patient during surgical or interventional procedures. GE has received one complaint of a transfer board not releasing from the MR table preventing the transfer board from moving to the surgical table. There have been no injuries reported as a result of this issue.