DIGITAL FLUOROGRAPHY SYSTEM, ULTIMAX-I & REMOTE CONTROL R/F SYSTEM ZEXIRA (2015-10-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Toshiba Medical Systems Corporation
1385 Shimoishigami
OTAWARA-SHI
324-8550
JAPAN
B. ULTIMAX-IDREX-U180_/K1
More than 10 numbers, contact manufacturer.
Manufacturer
Toshiba Medical Systems Corporation
1385 Shimoishigami
OTAWARA-SHI
324-8550
JAPAN
C. REMOTE CONTRL R/F SYSTEM ZEXIRA - MAIN UNITDREX-ZX80_/K4
More than 10 numbers, contact manufacturer.
Manufacturer
Toshiba Medical Systems Corporation
1385 Shimoishigami
OTAWARA-SHI
324-8550
JAPAN
A. DIGITAL FLUOROGRAPHY SYSTEMHDR-08A

Introduction

Affected Products
  1. DIGITAL FLUOROGRAPHY SYSTEM
  2. ULTIMAX-I
  3. REMOTE CONTRL R/F SYSTEM ZEXIRA - MAIN UNIT

Reason

By the inappropriate processing of the software for the image-processing device(HDR-08A), one-shot images are saved as black images in the hard disk, and they are displayed as black images on the system monitor, even though these images are displayed normally on the live monitor.