Health Products

NxStage System One-Cycler (Chronic Application) (2015-10-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
NxStage Medical, Inc.
350 Merrimack Street
Lawrence
01843
Massachusetts
UNITED STATES
NxStage System One-Cycler (Chronic Application)Not Applicable

Introduction

Affected Products

NxStage System One-Cycler (Chronic Application)

Reason

NxStage has identified a software error in the NxStage System One S Cycler Model No. Nx1000-3 with software versions 4.9 and 4.10. If specific conditions are met and a software error does occur, the ultrafiltration (UF) volume target will not decrease during the treatment even if the uf pump is running.