NxStage has identified a software error in the NxStage System One S Cycler Model No. Nx1000-3 with software versions 4.9 and 4.10. If specific conditions are met and a software error does occur, the ultrafiltration (UF) volume target will not decrease during the treatment even if the uf pump is running.