Health Products

YSIO MAX (2015-11-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
24102
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
YSIO MAX10762470

Introduction

Affected Products

YSIO MAX

Reason

Siemens has received complaints regarding the 3D tube head of the Ysio System moving to the planned and programmed end position faster than expected and then stopping with an error code. A system reboot is then required. This issue is due to a software bug.