MPS Delivery Set (2015-10-29)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
0492205E03
Manufacturer
Quest Medical, Inc.
ONE ALLENTOWN PARKWAY
Allen
75002
Texas
UNITED STATES
B. MPS Delivery Set5001102AS
More than 10 numbers, contact manufacturer.
Manufacturer
Quest Medical, Inc.
ONE ALLENTOWN PARKWAY
Allen
75002
Texas
UNITED STATES
A. MPS Delivery Set w/arrest agent and additive cassette5001102

Introduction

Affected Products

A. MPS Delivery Set w/arrest agent and additive cassette
B. MPS Delivery Set

Reason

The identified lots of MPS Delivery Sets have shown a possible seal failure along the blood source channel of main pump cassette, resulting in blood loss from the bypass circuit and interruption of cardioplegia solution delivery.