Health Products

MYLA SERVER HP PROLIANT DL380-G8 IVD MYLA SERVER (2015-11-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Myla Clinical 3.X to 4.1 DL380
Manufacturer
Biomerieux Inc.
595 Anglum Road
Hazelwood
63042
Missouri
UNITED STATES
MYLA SERVER HP PROLIANT DL380-G8 IVD MYLA SERVER415358 MASTER 3.2 CLI DL380

Introduction

Affected Products

MYLA SERVER HP PROLIANT DL380-G8 IVD MYLA SERVER

Reason

Under certain conditions, Myla Software, when connected to a Vitek 2 System and to a laboratory information system (LIS) allowing the re-use of specimen ID, presents the potential to link a result from Vitek 2 to the wrong patient and then upload those results to the LIS.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime