Health Products

BladderScan BVI 9600 and AortaScan AMI 9700 (2015-11-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Verathon Inc.
20001 NORTH CREEK PARKWAY
BOTHELL
98011
Washington
UNITED STATES
B. AortaScan AMI 9700AMI 9700 SYSTEM
All lots.
Manufacturer
Verathon Inc.
20001 NORTH CREEK PARKWAY
BOTHELL
98011
Washington
UNITED STATES
A. BladderScan BVI 9600BVI 9600 SYSTEM

Introduction

Affected Products

A. BladderScan BVI 9600
B. AortaScan AMI 9700

Reason

Notification that use of medical device was not intended to include detection of abdominal aortic aneurysms (AAA). A negative result may underreport the actual diameter of a patient's abdominal aorta and may result in a false negative AAA detection.