Health Products

Anorectal and Esophageal Manometric Catheters (2015-10-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
DS708
Manufacturer
Mui Scientific
H & A Mui Enterprises Inc.
145 Traders Boulevard East, Unit 34
Mississauga
L4Z 3L3
Ontario
CANADA
B. ESOPHAGEAL MANOMETRIC CATHETERSCE2-A-E27-LOSS-1-S
D31453
Manufacturer
Mui Scientific
H & A Mui Enterprises Inc.
145 Traders Boulevard East, Unit 34
Mississauga
L4Z 3L3
Ontario
CANADA
A. ANORECTAL MANOMETRIC CATHETERSCR4-XXXX

Introduction

Affected Products
  1. Anorectal Manometric Catheters
  2. Esophageal Manometric Catheters

Reason

Mui Scientific has removed an error from the instructions for use. Reference to high-level disinfection was removed as silicone catheters can only be sterilized (autoclaved) and not high-level disinfected.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime