Health Products

H3+ (PLUS) HOLTER RECORDER (2015-11-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Mortara Instrument Inc.,
7865 North 86th St.,
H3+ (PLUS) HOLTER RECORDERBURH3PR

Introduction

Affected Products

H3+ (PLUS) HOLTER RECORDER

Reason

An issue with the placement of a capacitor in the H3+ Digital Holter 48 hour recorders has resulted in reduced battery life, causing the recorder to record for less than the expected time. H3+ Digital Holter recorders with serial numbers between 115110248400 and 115410276012 fail to meet the claim of 24-48 hours recording time.