Health Products

COPPER NITI PREFORMED ARCHWIRE, 40C, .016X .022 LO PK 10 (2015-11-02)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
021538339
Manufacturer
Ormco Corporation
1332 South Lone Hill Avenue
Glendora
91740
California
UNITED STATES
COPPER NITI PREFORMED ARCHWIRE, 40C, .016X .022 LO PK 10211-0952

Introduction

Affected Products

COPPER NITI PREFORMED ARCHWIRE, 40C, .016X .022 LO PK 10

Reason

The product transformation temperature as listed on the labelling is not correct.  The wire is labelled as having a 40°C transformation temperature, but the actual wire has a transformation temperature of 29°C.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime