Two identified problems that Human Care has decided to address. The first problem is related to a fixation screw on the telescopic hanger bar used on recalled device. This screw is located inside the unit, hence not visible from the outside. A small number of units have been detected where the level of Loctite adhesive applied to the screw during manufacturing has been insufficient. This can cause the telescopic hanger bar to become lose from its position and, in a worst case, fall out of its position.
The second problem is related to the swivel hooks on the telescopic hanger bar. It has come to Human Care’s attention that if the telescopic hanger bar is incorrectly used, specifically if the swivel hooks on the hanger bar is positioned incorrectly before a patient transfer, this can cause the sling to fall out of its position which can then result in a hazard to the user. It is important that the user manual is followed and that the hanger bar and its hooks are adjusted and positioned correctly prior to any patient transfer.