ADVIA Centaur Assays (2015-11-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
73544070B. ADVIA CENTAUR CP SYS-HIV AG/AB COMBO (CHIV) ASSAY
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
06520528
73544070A. Advia Centaur SYS-HIV AG/AB Combo Assay
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
06520528

Introduction

Affected Products
  1. Advia Centaur SYS-HIV AG/AB Combo Assay
  2. ADVIA CENTAUR CP SYS-HIV AG/AB COMBO (CHIV) ASSAY

Reason

Siemens Healthcare Diagnostics has confirmed reports of calibration failures on the ADVIA Centaur HIV Ag/Ab Combo (CHIV) kit lots listed in section 1 due to imprecision of the low calibrator. An invalid calibration status will prevent CHIV results from being generated. Siemens has confirmed that due to sporadic aberrantly high relative light units (rlus), imprecision may be observed with the low calibrator, negative quality control (qc) material or negative samples.