Health Products

VITROS 4600 and 5,1 FS Chemistry Systems (2015-11-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
B. VITROS 5,1 FS Chemistry System680 1375
More than 10 numbers, contact manufacturer.
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
C. VITROS 4600 Chemistry System680 2445
More than 10 numbers, contact manufacturer.
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
A. VITROS 5,1 FS Chemistry System680 1375

Introduction

Affected Products
  1. VITROS 5,1 FS Chemistry System
  2. VITROS 5,1 FS Chemistry System
  3. VITROS 4600 Chemistry System

Reason

Ortho Clinical Diagnostics (OCD) has issued this recall due to a software anomaly that may potentially occur on Vitros 4600 or 5,1 FS Systems. OCD recently identified that under very specific conditions, the temperature in the Microwell Subsystem (included on Vitros 3600 and 5600 Systems) may potentially be out of range without the system properly alerting the operator. Although Vitros 4600 and 5,1 FS systems do not include a Microwell Subsystem, it utilizes the same software as Vitros 3600 and 5600 Systems. OCD has estimated that there is a less than 1 in 14 million possibility that this anomaly could occur in the Microslide or Microtip subsystems included on Vitros 4600 or 5,1 FS Systems.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime