ADVIA CENTAUR CP SYSTEM THYROXINE T4 ASSAY (2015-11-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
B. ADVIA CENTAUR SYSTEM - T4 ASSAY110735
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
A. ADVIA CENTAUR CP SYSTEM - THYROXINE (T4) ASSAY110735

Introduction

Affected Products
  1. ADVIA CENTAUR CP SYSTEM - THYROXINE (T4) ASSAY
  2. ADVIA CENTAUR SYSTEM - T4 ASSAY

Reason

Siemens Healthcare Diagnostics is currently investigating confirmed reports of calibration slope failures on the ADVIA Centaur Systems T4 Reagent Lots listed in section 64(a). A failed calibration slope will result in an invalid calibration status which will prevent T4 patient results from being generated. If the calibration slope passes and controls are within laboratory limits, patient results are considered valid and may be reported.