Health Products

FLEXIBLE SCREWDRIVER, BONE AWL & DRILL (2015-11-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots. 
Manufacturer
Aesculap AG & Co. KG
AESCULAP PLATZ
TUTTLINGEN
78532
GERMANY
B. FLEXIBLE BONE AWLSJ607R
All lots.
Manufacturer
Aesculap AG & Co. KG
AESCULAP PLATZ
TUTTLINGEN
78532
GERMANY
C. FLEXIBLE DRILLSJ723R
All lots.
Manufacturer
Aesculap AG & Co. KG
AESCULAP PLATZ
TUTTLINGEN
78532
GERMANY
A. FLEXIBLE SCREWDRIVER SJ706RSJ706R

Introduction

Affected Products
  1. FLEXIBLE SCREWDRIVER SJ706R
  2. FLEXIBLE BONE AWL
  3. FLEXIBLE DRILL

Reason

The mechanical strength of the flexible shaft does not reliably meet the requirements of surgery. This leads to forced brittle fracture under bending loads causing fragments to be left behind.