Health Products

IMPAX (2015-11-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
IMPAX 6.5.x and above
Manufacturer
AGFA Healthcare N.V.
Septestraat 27
Mortsel
2640
BELGIUM
A. IMPAX6.X

Introduction

Recalled Products

A. IMPAX

Reason

When using the extended addendum workflow in impax 6.5.x and above with embedded reporting there is potential for loss of report text when validating addendums. This occurs when all of the conditions listed below happen at the same time:

  • the dictation or report text is not started, as intended, at the initial cursor location, and


  • the keyboard arrow keys, mouse or SpeechMike input are used to move the cursor in the addendum before dictating or typing report text, and


  • the "REPORTSTATUS" bookmark is used to automatically add the status to the addendum, and


  • with respect to Impax 6.5.4 and prior 6.5.x versions, the "BEGINPROTECT" and "ENDPROTECT" bookmarks have not been used to protect the "REPORTSTATUS" bookmark.
If all of the above conditions are met, and the cursor is placed at the "REPORTSTATUS" bookmark location when starting to dictate or type report text, Impax will overwrite the text entered at the "REPORTSTATUS" bookmark location with the addendum status when the addendum is validated. As a result, some of the text of the addendum is not saved in the validated addendum.