Health Products

TUBE SPC EDTA PLH 13X75 3.0 PLBL TAN SP (2015-12-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
5064684
Manufacturer
Becton Dickinson and Company
1 Becton Drive
Franklin Lakes
07417
New Jersey
UNITED STATES
A. TUBE SPC EDTA PLH 13X75 3.0 PLBL TAN SP367855

Introduction

Recalled Products

A. TUBE SPC EDTA PLH 13X75 3.0 PLBL TAN SP

Reason

The product has the potential to contain the incorrect (ethylenediaminetetraacetic acid) EDTA additive volume. Various amounts of EDTA additive in the tubes may lead to platelet clumping and clotting of the blood sample. Platelet clumping causes false reduction of platelet counts, the possible consequences of false reduction of platelet counts are erroneous results and specimen rejection. In the event that the clinician acts upon the erroneous results, this could lead to unnecessary platelet transfusion, hematologic work up including bone marrow biopsy, steroid therapy, splenectomy etc.