Recalled Products
A. TUBE SPC EDTA PLH 13X75 3.0 PLBL TAN SP
The product has the potential to contain the incorrect (ethylenediaminetetraacetic acid) EDTA additive volume. Various amounts of EDTA additive in the tubes may lead to platelet clumping and clotting of the blood sample. Platelet clumping causes false reduction of platelet counts, the possible consequences of false reduction of platelet counts are erroneous results and specimen rejection. In the event that the clinician acts upon the erroneous results, this could lead to unnecessary platelet transfusion, hematologic work up including bone marrow biopsy, steroid therapy, splenectomy etc.