UNICEL DXH COULTER ANALYSIS SYSTEM (2015-12-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots.
Manufacturer
Beckman Coulter, Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
B. UNICEL DXH SLIDEMAKER STAINER COULTER CELLULAR ANALYSIS SYSTEM775222
All lots.
Manufacturer
Beckman Coulter, Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
C. UNICEL DXH 600 COULTER CELLULAR ANALYSIS SYSTEMB23858
All lots.
Manufacturer
Beckman Coulter, Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
A. UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM629029

Introduction

Affected Products
  • A. UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
  • B. UNICEL DXH SLIDEMAKER STAINER COULTER CELLULAR ANALYSIS SYSTEM
  • C. UNICEL DXH 600 COULTER CELLULAR ANALYSIS SYSTEM

Reason

Beckman Coulter determined that the affected software for the DXH systems allows the creation of multiple orders with the same specimen identification (ID) but different patient identification when manually editing pending orders at the system manager. The issue does not occur for edits of released results or for test orders requested through host transmission. The issue creates the potential for sample misidentification and potential of releasing erroneous results.