Health Products

m2000sp Instrument, G-series and E-series (2015-12-14)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
10378
Manufacturer
Abbott Molecular Inc.
1300 E. Touhy Ave.
Des Plaines
60018
Illinois
UNITED STATES
A. m2000sp Instrument, G-series and E-series9K14-01

Introduction

Affected Products

A. m2000sp Instrument, G-series and E-series

Reason

Investigation at Abbott Molecular identified that an incorrect type of tip adapter (type D) could have been installed on m2000sp instrument during a service visit on or after august 14, 2015. This is because two lots of Abbott Molecular PN 30084426 (lots 76258007 and 77228007) were shipped with type D tip adapters rather than the required type B adapters.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime