Health Products

Dialog+ SW 9.XX (2015-11-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lots.A. Dialog+ SW 9.XX
Manufacturer
B. Braun Avitum Ag
Schwarzenberger Weg 73-79
Melsungen
34212
GERMANY
710200CV

Introduction

Affected Products

A. Dialog+ SW 9.XX

Reason

Potential problem if servicing device and interchanging tubing between valve block level regulation and diaphragm pump if the functionality of the level regulation system is not tested as requested in the service manual. An interchanged tubing results in an inverse function of the manual level regulation system.