B. ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING - PIP/TAZO/CON-4
412434
1002092270
Manufacturer
Biomerieux Inc.
595 Anglum Road
Hazelwood
63042
Missouri
UNITED STATES
A. ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING - PIPERACILLIN/TAZOBACTAM
521418
Introduction
Affected Products
A. ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING - PIPERACILLIN/TAZOBACTAM
B. ETEST ANTIMICROBIAL SUSCEPTIBILITY TESTING - PIP/TAZO/CON-4
Reason
On October 27th, 2015, a customer reported a low Mic on Etest Piperacillin/Tazobactam (4 ug/ml) PTC-256 single pack (lot 1004271120). The customer obtained erroneous results (false susceptible) on Etest with broth-micro dilution reference method.