Manufacturer
Zimmer Inc.
1800 West Center Street
Warsaw
46580
UNITED STATES
A. CUSTOM TRILOGY STYLE ALLOFIT IT INSERTER82-0174-140-00

Introduction

Affected Products

A. CUSTOM TRILOGY STYLE ALLOFIT IT INSERTER

Reason

Zimmer Biomet is initiating a lot specific recall of the Trilogy style Allofit it inserters, part number 82-0174-140-00. Zimmer Biomet received two complaints for three devices which had bent and cracked at the weld which connects the Allofit it inserter handle to the Allofit it inserter tip. The affected devices were distributed between 25 June 2011 and 28 December 2012.