Health Products

ENTERALITE INFINITY ENTERAL FEEDING SETS (2015-12-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lots.B. Enteralite Infinity Spike Set
Manufacturer
Zevex Inc.
also trading as Moog Medical Devices Group
4314 Zevex Park Lane
Salt Lake City
84123
Utah
UNITED STATES
INF0020-A
All lots.A. ENTERALITE INFINITY BAG SET
Manufacturer
Zevex Inc.
also trading as Moog Medical Devices Group
4314 Zevex Park Lane
Salt Lake City
84123
Utah
UNITED STATES
INF0500-A

Introduction

Affected Products

A. ENTERALITE INFINITY BAG SET
B. Enteralite Infinity Spike Set

Reason

The manufacturer has become aware of an increase in the number of complaints for EnterLite Infinity Enteral Pump Delivery sets with ENFit connector system and transitional stepped connectors (administration sets) that are used with the EnteraLite Infinity Enteral Feeding Pump.  Customers are reporting formula leaking between the purple ENFit connector and the white transitional stepped connector due to cracking or breaking of the purple connector.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime