Health Products

ALLURA XPER SYSTEMS (2015-12-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
55114616
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6, 5684 PC
Best, Netherlands
F. ALLURA XPER FD20 OR TABLE000454
More than 10 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6, 5684 PC
Best, Netherlands
D. ALLURA XPER FD10ALLURA XPER FD10
More than 10 numbers, contact manufacturer
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6, 5684 PC
Best, Netherlands
E. ALLURA XPER FD20 SYSTEM - MAIN UNITALLURA XPER FD20
53388068
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6, 5684 PC
Best, Netherlands
B. ALLURA XPER FD20/10 SYSTEM - MAIN UNITALLURA XPER FD20/10
53329363
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6, 5684 PC
Best, Netherlands
C. ALLURA XPER FD20/20ALLURA XPER FD20/20
53772614
Manufacturer
Philips Medical Systems Nederland B.V.
Veenpluis 4-6, 5684 PC
Best, Netherlands
A. ALLURA XPER FD10/10 - MAIN UNITALLURA XPER FD10/10

Introduction

Affected Products

A. ALLURA XPER FD10/10 - MAIN UNIT
B. ALLURA XPER FD20/10 SYSTEM - MAIN UNIT
C. ALLURA XPER FD20/20
D. ALLURA XPER FD10
E. ALLURA XPER FD20 SYSTEM - MAIN UNIT
F. ALLURA XPER FD20 OR TABLE

Reason

Philips has discovered through customer feedback and internal testing a failure with the image disk causing the entire image processing chain to be blocked due to a software design error. This causes image data to be lost, and eventually loss of live imaging functionality
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime