Health Products

LIFECODES HLA-DQA1/B1 Typing Kit (2015-12-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
07065BA. LIFECODES HLA-DQA1/B1 Typing Kit
Manufacturer
Immucor Transplant Diagnostics, Inc.
550 West Avenue
Stamford
06902
Connecticut
UNITED STATES
628930

Introduction

Affected Products

A. LIFECODES HLA-DQA1/B1 Typing Kit

Reason

Immucor Transplant Diagnostics Inc., is investigating a single instance of false positive reactivity of probe dqb253 associated with an allele pair of DQB1*03:03:02:01, DQB1*03:04:01.  A false positive result causes the exclusion of DQB1*03:03:02:01/ 03:04:01 as a possible allele combination.  The false positive reactivity is only observed in a DQB1*03:03:02:01, DQB1*03:04:01 heterozygous combination. Other combinations with DQB1*03:03:02:01 or DQB1*03:04:01 are not affected.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime