Health Products

OPTIMA XR200AMX X-RAY SYSTEM - MAIN UNIT (2015-12-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
GE Medical Systems LLC
3200 North Grandview Blvd
Waukesha
53188
Wisconsin
UNITED STATES
A. OPTIMA XR200AMX X-RAY SYSTEM - MAIN UNIT5416437GTC

Introduction

Affected Products

A. OPTIMA XR200AMX X-RAY SYSTEM - MAIN UNIT

Reason

GE Healthcare has recently become aware of an issue related to the operation of the safety latch within the column that supports the horizontal arm. This safety latch mechanism serves as a backup in the unlikely event the primary safety cable breaks. A fall of the arm assembly could result in bodily harm to a person. There have been no incidents or injuries reported as a result of this internally found issue.