Health Products

ONCENTRA EXTERNAL BEAM/ONCENTRA BRACHY 4.5 (2015-12-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
19311-EB001
Manufacturer
Nucletron B.V.
Waardgelder 1
Veenedaal
3905 TH
NETHERLANDS
A. ONCENTRA EXTERNAL BEAM/ONCENTRA BRACHY 4.5170.730

Introduction

Affected Products

A. ONCENTRA EXTERNAL BEAM/ONCENTRA BRACHY 4.5

Reason

When using the option "Tumor Overlap Fraction" in VMAT planning it has been observed that in rare cases the system does include an organ at risk as target volume. This could result in open MLC, and open jaws in areas away from the target volume.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime