Symbiq Infusion System : Pole Clamp Assembly (2015-12-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All serial numbers.A. Symbiq Infusion System : Pole Clamp Assembly
Manufacturer
Hospira Inc.
275 North Field Drive
Lake Forest
60045
Illinois
UNITED STATES
16026; 16027                                                                                        

Introduction

Affected Products

A. Symbiq Infusion System : Pole Clamp Assembly

Reason

SYMBIQ pole clamp assemblies used to repair the Symbiq Infuser have been identified to be missing the rubber stop pad. This pad reduces the potential for the Symbiq infusion pump to slip down an IV pole.  All serial numbers are potentially affected as they all have the potential for an affected pole clamp assembly to be installed on them during repair. The part number for the impacted pole clamp assembly is HSP3313-E12 the affected lot # is 481976