Health Products

CARDIOSAVE SYSTEM - HYBRID & RESCUE INTRA-AORTIC BALLOON PUMP (2015-12-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Datascope Corp.
1300 Macarthur Blvd.
Mahwah
07430
New Jersey
UNITED STATES
A. CARDIOSAVE SYSTEM - HYBRID & RESCUE INTRA-AORTIC BALLOON PUMP0998-00-0800-XX

Introduction

Affected Products
  1. CARDIOSAVE SYSTEM - HYBRID & RESCUE INTRA-AORTIC BALLOON PUMP 

Reason

In some CARDIOSAVE IABPs, the scroll compressor did not meet the specifications for output pressure or vacuum at specific flow rates. When the scroll compressor fails, one of two high priority alarms will appear on the IABP display and patient therapy could be interrupted.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime