A. CARDIOSAVE SYSTEM - HYBRID & RESCUE INTRA-AORTIC BALLOON PUMP
0998-00-0800-XX
Introduction
Affected Products
CARDIOSAVE SYSTEM - HYBRID & RESCUE INTRA-AORTIC BALLOON PUMP
Reason
In some CARDIOSAVE IABPs, the scroll compressor did not meet the specifications for output pressure or vacuum at specific flow rates. When the scroll compressor fails, one of two high priority alarms will appear on the IABP display and patient therapy could be interrupted.