Toshiba Medical Systems Corporation
1385 Shimoishigami
OTAWARA-SHI
324-8550
JAPAN
A. RADREX-IDRAD-3000E_/W8

Introduction

Affected Products
  1. RADREX-I

Reason

When an operator performs radiography with Wireless FPD, the following message may be displayed on the monitor: "The image cannot be acquired from FPD. Press ok to try retrieving." however, when the operator selects "ok" or "cancel" on the message window, the image is not acquired and is lost.