Datex-Ohmeda Anesthesia Delivery System Main Units (2015-12-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Datex-Ohmeda
3030 Ohmeda Drive, P.O. Box 7550,
Madison
53707-7550
Wisconsin
UNITED STATES
B. AVANCE CS2 - Main Unit1009-9050-000
More than 10 numbers, contact manufacturer
Manufacturer
Datex-Ohmeda
3030 Ohmeda Drive, P.O. Box 7550,
Madison
53707-7550
Wisconsin
UNITED STATES
C. AISYS CS2 ANESTHESIA MACHINE MAIN UNIT1011-9050-000
More than 10 numbers, contact manufacturer
Manufacturer
Datex-Ohmeda
3030 Ohmeda Drive, P.O. Box 7550,
Madison
53707-7550
Wisconsin
UNITED STATES
A. S/5 AESPIRE 7900 SMARTVENT ANESTHESIA SYSTEM - Main Unit1009-9012-000

Introduction

Affected Products
  1. S/5 AESPIRE 7900 SMARTVENT ANESTHESIA SYSTEM - Main Unit
  2. AVANCE CS2 - Main Unit
  3. AISYS CS2 ANESTHESIA MACHINE MAIN UNIT

Reason

The drive gas check valve could become stuck in a fixed open position which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma.