B. ADVIA CENTAUR SYSTEM - PROSTATE-SPECIFIC ANTIGEN
118157
ending in 263 and above
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
A. ADVIA CENTAUR CP SYSTEM - PROSTATE-SPECIFIC ANTIGEN
118156
Introduction
Affected Products
ADVIA CENTAUR CP SYSTEM - PROSTATE-SPECIFIC ANTIGEN
ADVIA CENTAUR SYSTEM - PROSTATE-SPECIFIC ANTIGEN
Reason
Not meeting the current High-Dose Hook Effect expectation in the Instructions For Use (IFU). The ADVIA Centaur systems PSA assay range is 0.01-100 ng/mL (μg/L). The ADVIA Centaur Systems PSA IFU states that patient samples with high total PSA levels can cause a paradoxical decrease in the Relative Light Units (RLUs) (high-dose hook effect). In this assay, patient samples with total PSA levels as high as 10,000 ng/mL (μg /L) will assay greater than 100 ng/mL (μg/L).