ALPHA-METHYLACYL-COA RACEMASE (AMACR P504S) and BONDS READY-TO-USE (2015-12-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
all lots 
Manufacturer
Leica Biosystems Newcastle Ltd.
Balliol Business Park West
Benton Lane
UNITED KINGDOM
H. BOND READY-TO-USE PRIMARY ANTIBODY ALPHA-METHYLACYL-COA RACEMASE (EPMU1)PA0210
all lots
Manufacturer
Leica Biosystems Newcastle Ltd.
Balliol Business Park West
Benton Lane
UNITED KINGDOM
F. BOND READY-TO-USE PRIMARY ANTIBODY PAX-5 (1EW)PA0552
all lots 
Manufacturer
Leica Biosystems Newcastle Ltd.
Balliol Business Park West
Benton Lane
UNITED KINGDOM
G. BOND READY-TO-USE ISH CMV PROBEPB0614
all lots 
Manufacturer
Leica Biosystems Newcastle Ltd.
Balliol Business Park West
Benton Lane
UNITED KINGDOM
D. BOND READY-TO-USE ISH LAMBDA PROBEPB0669
all lots
Manufacturer
Leica Biosystems Newcastle Ltd.
Balliol Business Park West
Benton Lane
UNITED KINGDOM
E. BOND READY-TO-USE ISH RNA CONTROL PROBEPB0785
all lots
Manufacturer
Leica Biosystems Newcastle Ltd.
Balliol Business Park West
Benton Lane
UNITED KINGDOM
B. BOND READY-TO-USE ISH EBER PROPB0589
all lots
Manufacturer
Leica Biosystems Newcastle Ltd.
Balliol Business Park West
Benton Lane
UNITED KINGDOM
C. BOND READY-TO-USE ISH KAPPA PROBEPB0645
all lots
Manufacturer
Leica Biosystems Newcastle Ltd.
Balliol Business Park West
Benton Lane
UNITED KINGDOM
A. NOVOCASTRA LIQUID MOUSE MONOCLONAL ANTIBODY ALPHA-METHYLACYL-COA RACEMASE (AMACR, P504S)NCL-L-AMACR

Introduction

Affected products
  1. NOVOCASTRA LIQUID MOUSE MONOCLONAL ANTIBODY ALPHA-METHYLACYL-COA RACEMASE (AMACR, P504S)
  2. BOND READY-TO-USE ISH EBER PRO
  3. BOND READY-TO-USE ISH KAPPA PROBE
  4. BOND READY-TO-USE ISH LAMBDA PROBE
  5. BOND READY-TO-USE ISH RNA CONTROL PROBE
  6. BOND READY-TO-USE PRIMARY ANTIBODY PAX-5 (1EW)
  7. BOND READY-TO-USE ISH CMV PROBE
  8. BOND READY-TO-USE PRIMARY ANTIBODY ALPHA-METHYLACYL-COA RACEMASE (EPMU1)

Reason

These antibodies and probes, when used in combination with specified lots of the Bond Polymer Refine Detection and Novolink Polymer Detection System, may not provide adequate staining when used according to the instructions for use. If the products are used according to their instructions for use with appropriate positive controls, the failure to adequately stain would be apparent to the user on the control tissue and on the patient tissue. The absence of adequate staining of tissue is likely to result in a delay in obtaining the results of IHC / ISH staining and, in a worst-case scenario, could cause a delay in the diagnosis or classification of a neoplasm.