Ambulatory Infusion Pumps (2015-12-24) Dec 24, 2015 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number N/A Manufacturer Zevex Inc. Also Trading as Moog Medical Devices Group 4314 Zevex Park Lane Salt Lake City Utah UNITED STATES B. Curlin CMS 4000 Ambulatory Volumetric Infusion Pump 3501000 N/A Manufacturer Zevex Inc. Also Trading as Moog Medical Devices Group 4314 Zevex Park Lane Salt Lake City Utah UNITED STATES C. PAINSMART IOD (AMBULATORY INFUSION PUMP) N/A Manufacturer Zevex Inc. Also Trading as Moog Medical Devices Group 4314 Zevex Park Lane Salt Lake City Utah UNITED STATES A. Curlin CMS 6000 Ambulatory Volumetric Infusion Pump 3601100
Affected Products
Curlin CMS 6000 Ambulatory Volumetric Infusion Pump
Curlin CMS 4000 Ambulatory Volumetric Infusion Pump
PAINSMART IOD (AMBULATORY INFUSION PUMP)
The master pumps used to calibrate manufactured and serviced pumps have a calibration shift of up to 6.8% (+3%). When an affected master pump is used to calibrate production and serviced pumps, this pump is used to calibrate production and serviced pumps, this calibration shift could lead to over delivery of fluid beyond the label claim of +5%.