Health Products

Ambulatory Infusion Pumps (2015-12-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
N/A
Manufacturer
Zevex Inc. Also Trading as Moog Medical Devices Group
4314 Zevex Park Lane
Salt Lake City
Utah
UNITED STATES
B. Curlin CMS 4000 Ambulatory Volumetric Infusion Pump3501000
N/A
Manufacturer
Zevex Inc. Also Trading as Moog Medical Devices Group
4314 Zevex Park Lane
Salt Lake City
Utah
UNITED STATES
C. PAINSMART IOD (AMBULATORY INFUSION PUMP)
N/A
Manufacturer
Zevex Inc. Also Trading as Moog Medical Devices Group
4314 Zevex Park Lane
Salt Lake City
Utah
UNITED STATES
A. Curlin CMS 6000 Ambulatory Volumetric Infusion Pump3601100

Introduction

Affected Products
  1. Curlin CMS 6000 Ambulatory Volumetric Infusion Pump
  2. Curlin CMS 4000 Ambulatory Volumetric Infusion Pump
  3. PAINSMART IOD (AMBULATORY INFUSION PUMP)

Reason

The master pumps used to calibrate manufactured and serviced pumps have a calibration shift of up to 6.8% (+3%). When an affected master pump is used to calibrate production and serviced pumps, this  pump is used to calibrate production and serviced pumps, this calibration shift could lead to over delivery of fluid beyond the label claim of +5%.