Health Products

EliA SmDp Well (2015-12-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
18B. EliA SmDp Well - PHADIA 100
Manufacturer
PHADIA AB
RAPSGATAN 7
UPPSALA
SWEDEN
14-5624-01
18A. EliA SmDp Well - PHADIA 250
Manufacturer
PHADIA AB
RAPSGATAN 7
UPPSALA
SWEDEN
14-5624-01

Introduction

Affected Products
  1. EliA SmDp Well - PHADIA 250
  2. EliA SmDp Well - PHADIA 100

Reason

Analysis of patient samples received in complaint investigations showed that several samples caused unspecific signals up to 22 U/mL. The unspecific signals are neither caused by anti-Sm antibodies nor by anti-streptavidin antibodies but are due to other unspecific signals with an as yet unknown cause.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime