Medrad Intego Pet Infusion System - Source Admini (2015-11-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 lots, contact manufacturer
Manufacturer
Bayer Medical Care Inc.
1 Bayer Drive
Indianola
15051-0780
Pennsylvania
UNITED STATES
MEDRAD INTEGO PET INFUSION SYSTEM - SOURCE ADMINIINT CSS

Introduction

Affected Products

MEDRAD INTEGO PET INFUSION SYSTEM - SOURCE ADMINI

Reason

The inner diameter of the saline tubing is out of specification and may result in the system failing to position the dose correctly during extraction, potentially leading to a low measurement of the extracted dose or to a higher dose being extracted. The occluded SAS tubing may impair the priming function and prompt an "RP prime failed. Check SAS and vial installation, then reprime" error message.