Keypoint System - EP Amplifier (Headbox) (2015-11-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Version 2.31 - 2.32 software
Manufacturer
Alpine Biomed Aps
Tonsbakken 16-18
Skovlunde
KEYPOINT SYSTEM - EP AMPLIFIER (HEADBOX)9031C077_

Introduction

Affected Products

KEYPOINT SYSTEM - EP AMPLIFIER (HEADBOX)

Reason

The channel configuration in the EP amplifier is not always fully restored when returning to the waveform display. Intermittently the effected channel (s) becomes disconnected from the EP amplifier input pins, leaving the input signal floating. The results in no curve displayed or the displayed curve (s) becoming distorted (noisy, unstable).