Health Products

Astral 150 - AMER Astral 150 - FRA (2016-01-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Resmed Limited
1 Elizabeth Macarthur Drive
Bella Vista
2153
AUSTRALIA
Astral 150 - AMER Astral 150 - FRA27003

Introduction

Affected Products
  1. Astral 150 - AMER Astral 150 - FRA

Reason

There have been nine complaints globally of the ventilator entering non-responsive state while connected to an external battery system which can cause a delay in treatment. This event can occur after the device is in a standby state only and does not occur during therapy. This only affects devices with Astral Service Release 1.1 (SR 1.1) software.