ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEMS AND STRIPS (2016-01-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots 
Manufacturer
Iris Diagnostics, a Division of Iris International Inc.
9172 Eton Avenue
Chatsworth
UNITED STATES
B. ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM W LOAD/UNLOAD800-7167
All Lots
Manufacturer
Iris Diagnostics, a Division of Iris International Inc.
9172 Eton Avenue
Chatsworth
UNITED STATES
C. ICHEM VELOCITY URINE CHEMISTRY STRIPS800-7212
All lots
Manufacturer
Iris Diagnostics, a Division of Iris International Inc.
9172 Eton Avenue
Chatsworth
UNITED STATES
ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM800-7719

Introduction

Affected Products
  1. ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM
  2. ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM W LOAD/UNLOAD
  3. ICHEM VELOCITY URINE CHEMISTRY STRIPS

Reason

IRIS international has become aware of loose or missing analyte pads being found in several locations (strip vial, strip provider module, strip conveyor system or waste container) that are undetectable by the iChem VELOCITY system. If affected strips are used to assay patient samples, the issue has the potential to cause erroneous pH results and false negative results for the following tests: nitrite, glucose, blood, bilirubin, urobilinogen, protein, leukocyte esterase, ketones, and ascorbic acid.