Health Products

BEAVER 4.0mm BANANA BLADE (2016-01-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Smith & Nephew, Inc., Endoscopy Division
150 Minuteman Road
Andover
01810
Massachusetts
UNITED STATES
BEAVER 4.0mm BANANA BLADE72203307

Introduction

Affected Products
  1. BEAVER 4.0mm BANANA BLADE

Reason

Several lots of the BEAVER 4.0mm banana blade are experiencing brittle fracture. A number of product complaints were received with this indication noted during surgery. In the event that the blade breaks and falls into the patient, it could potentially be difficult to retrieve and may lead to a surgical delay.