A. TRILOGY 100 VENTILATOR-CANADA
B. TRILOGY 100 VENTILATOR-INTERNATIONAL
C. TRILOGY 202, INTERNATIONAL
D. TRILOGY 202, DOMESTIC
E. TRILOGY 200, CANADA
Reason
Trilogy Ventilators with software versions 13.2.04, 13.2.05, 14.0.00 or 14.1.01 are susceptible to an inadvertent change between prescriptions. This may occur under a particular set of operating conditions and after a specific sequence of key presses without requiring confirmation of the change by the user. Affected devices continue to accurately display the active prescription in the upper left hand corner of the display. As this prescription change was unintended and no confirmation was required, the user may be unaware of this change.