ADVIA (1200) CHEMISTRY SYSTEM- GAMMA-GLUTAMYL TRANSFERASE (GGT) ASSAY (2016-01-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
360646
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
B. ADVIA 1200 CHEMISTRY SYSTEM -GAMMA-GLUTAMYL TRANSFERASE (GGT) ASSAY7498649
360646
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
A. ADVIA CHEMISTRY SYSTEM -GAMMA-GLUTAMYL TRANSFERASE (GGT) ASSAY7498649

Introduction

Affected Products

A. ADVIA CHEMISTRY SYSTEM -GAMMA-GLUTAMYL TRANSFERASE (GGT) ASSAY
B. ADVIA 1200 CHEMISTRY SYSTEM -GAMMA-GLUTAMYL TRANSFERASE (GGT) ASSAY

Reason

A portion of Gamma-Glutamyl Transferase Reagent Kit lot number 360646 for use on the ADVIA 1200, 1800, 2400 and XPT Chemistry systems may contain (R1) reagent wedges that have incorrectly printed and encoded barcode information. When placed on board the system, the affected R1 reagent wedge is not readable and is not recognized by the system.
Note: reagent containers that have the correct barcode can be used as is on the system. (R2 reagent has the correct barcode) 
correct R1 barcode: 740442164612736439 
incorrect R1 barcode: 747442164612736439