TERUMO CUVETTE (2016-01-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer. 
Manufacturer
Terumo Cardiovascular System Corporation
125 Blue Ball Road
Elkton
21921
Maryland
UNITED STATES
A. TERUMO CUVETTE6912

Introduction

Affected Products     

A. TERUMO CUVETTE 

Reason

CDI system 500 monitor displaying the "H/S DISCONNECT AT CUVETTE" error message when the CDI H/S cuvette does not make a proper connection to the CDI H/S probe. In this situation, blood parameter values for Hct, Hgb, and SO2 do not display.