GEHC has updated the service manuals used to maintain these scanners to fix an incorrect torque value, which is the source of the field action. FMI 25463 identified a torque issue with 5 volt supplies in CT and PET LightSpeed systems and corrected the issue by field action, as well as a technical manual update. This recall is to send a second customer letter to anyone who may still possess and use an obsoleted service manual.